Product Description
McLeod vacuum gauge(also called rotary compression vacuum gauge),
is based on the ideal isothermal compression Boyle - Mariotte law
designed. Because it can adopt its own parameters, pressure values
directly considered, it addition to the pressure measurement, but
also can be used as an absolute vacuum gauge vacuum gauge
calibration some relatively. It is widely used in industry,
national defense, scientific research, medicine, refrigeration and
other vacuum measurement work.
1、Before use, first picked up the vacuum exhaust pipe up and let
the vacuum table in the following, in order that the glass tube in
the process of mercury beads into the exhaust pipe. Then pull the
rubber plug, painted on the inner wall of the pipe mouth, castor
oil or vacuum grease is directly connected to the vacuum system to
be tested on the interface of open mechanical vacuum pump for 3-5
minutes, then slowly open the connection vacuum off-balance-sheet
vacuum valve, then the instrument enter into the state of work.
2、If the vacuum degree of vacuum system for rough vacuum, for PM -
type 2, then just look on the right side of the u-shaped pressure
vacuum gauge, the reading value is to look at the scale of bead
referred to value the sum of two mercury, when the vacuum degree is
less than 0.2 KPa (200 pa), two mercury end basic in parallel, it's
on the left side of the vacuum gauge, its operation method is the
same as the original McIntosh vacuum gauge (surface printing
office).
1、Transport and use must prevent inverted supine to prevent mercury
back into the vacuum hose or water inside the closed tube silver
beads into the middle clogged capillary capillary affect
measurement accuracy. Therefore, prior to each use must first
upright vacuum hose to the mercury can flow back into the vacuum
gauge glass tube.
2、When measuring the vacuum gauge for the upright position,it is
strictly prohibited to any gas strongly into the vacuum system,in
order to avoid shock tube shatter glass tube.
3、Mercury vacuum gauge has been working within the chemical physics
approach, a certain cleanliness. Long-term use due to reasons such
as pollution caused by mercury oxidation will directly affect the
measurement accuracy, should be promptly sent to repair, after
passing inspection before use.
4、After the vacuum meter in the measurement, application clip will
vacuum and vacuum system of the use of rubber tube clamping gas
separation. Clean, this can maintain a vacuum gauge in mercury can
also avoid the gage is removed separation, atmospheric strong
impact vacuum gauge so that it is broken. If not to get off vacuum
gauge, should after the mechanical vacuum pump stop, will be sure
to gas circuit disconnect mechanical pump and the vacuum system,
and put into the atmosphere in the mechanical pump in order to
avoid mechanical pump oil back into the vacuum system and vacuum
gauge polluted by oil.
Country: |
China |
Model No: |
-
|
FOB Price: |
(Negotiable)
Get Latest Price
|
Place of Origin: |
- |
Price for Minimum Order: |
- |
Minimum Order Quantity: |
1 Set |
Packaging Detail: |
Standard Packing |
Delivery Time: |
7 days after confirm the order |
Supplying Ability: |
5000 Set per Month |
Payment Type: |
Western Union, L/C, T/T |
Product Group : |
Measurement & Analysis Instruments
|